Guardian® Class II Biological Safety Cabinets
The Guardian® Guardian® Class II Biological Safety Cabinet
utilises a specially ventilated enclosure, developed for sterile
material handling.
The unit guarantees both operator and working materials protection from potential biohazard infections and complies to BS EN 12469:2000 and independently type tested by Public Health England.
Standard Features include:
- Low Energy - 120watts of power.
- UV Lighting Strip ensures VHP & Formalin Disinfection.
- Electric Sash Adjustment & Closure.
- H14 HEPA Filters that help create an ISO Class 4 Clean Environment.
- Air Diffuser and Sloped Intake for efficient and smooth airflow.
- Removable Polished Stainless Steel Sectional Work Surface.
- Infra-Red Intelligent Sash Safety (IRISS).
Visionaire® Touchscreen Control Interface
The Visionaire® Touchscreen Control Interface is designed to control & monitor all aspects of product operation including; Air Velocity (m/sec or ft/min), Sash Adjustment, UV Sterilisation Cycles, Electrical Sockets, Cabinet Power and Light Variable intensity.
Audible and Visual alarms alert the user to maintenance requirements including: Service, UV Light replacement and Cabinet Hours Run. Formalin and VHP Cycles are also controllable.
Size Specification (W x D x H)
MSC 800:
800mm x 750mm x 1351mm *
MSC 1200:
1200mm x 750mm x 1351mm *
MSC 1800:
1800mm x 750mm x 1351mm *
* +46 mm height for Outlet HEPA Filter.
Guardian ® Class I Biological Safety Cabinets
The Guardian® Class I Biological Safety Cabinet offers operator and environmental protection from biological materials and activities involving animal pathogens and genetically modified organisms.
Designed to incorporate the latest features to BS EN 12469:2000 Standards, the range of Microbiological Safety Cabinets offer total practicality and suitability for a huge variety of applications.
The units are provided as a re-circulating, or optionally ducted, system and utilise HEPA filtration to offer environmental and operator protection.
The H14 HEPAs are rated at 99.997% efficient at 0.3μm particulate, the filter seals under negative pressure to prevent leaks into the cabinet.
The MSC1200C1 can be fitted with Carbon Filters to allow use with a multitude of chemical compounds, including formalin. Universal ports allow connection to proprietary VHP and formalin systems to ensure safe and complete disinfection.
Size Specification (W x D x H)
MSC 1200 C1:
1200mm x 700mm x 1350mm *
*Add 15 mm height for Outlet Carbon Filter, if fitted.